Proceedings of the International School of Physics "Enrico Fermi.", Volume 76N. Zanichelli, 1981 - Nuclear physics |
From inside the book
Results 1-3 of 71
Page 126
... tissues . The term microcirculation is restricted arbitrarily to those subdivisions of the vascular tree that have diameters of less than 50 μm . These small vessels are not simply conduits , but form an integral part of the tissues ...
... tissues . The term microcirculation is restricted arbitrarily to those subdivisions of the vascular tree that have diameters of less than 50 μm . These small vessels are not simply conduits , but form an integral part of the tissues ...
Page 279
... tissues , carbon dioxide from the tissues to the lungs and hydrogen ions from the tissues to the kidneys . The function of the lungs is to execute the exchange of gases between the blood and the outside atmos- phere . Respiration is ...
... tissues , carbon dioxide from the tissues to the lungs and hydrogen ions from the tissues to the kidneys . The function of the lungs is to execute the exchange of gases between the blood and the outside atmos- phere . Respiration is ...
Page 405
... tissues are grouped together regardless of species . Moreover , ac- count is not taken either of tissue temperature or « freshness » . It is apparent from fig . 5 that , for soft tissues in the frequency range ( 0.1 ÷ 50 ) MHz , ( 2.12 ) ...
... tissues are grouped together regardless of species . Moreover , ac- count is not taken either of tissue temperature or « freshness » . It is apparent from fig . 5 that , for soft tissues in the frequency range ( 0.1 ÷ 50 ) MHz , ( 2.12 ) ...
Contents
A R D THORNTON | 1 |
Physicists and clinicians | 5 |
The Fourier transform properties of an image | 12 |
Copyright | |
48 other sections not shown
Other editions - View all
Common terms and phrases
alveolar amplitude analysis aorta aortic approximately arterial pressure arterial system attenuation value basilar membrane beam blood flow blood pressure C₁ capillary cardiac output circulation clinical cm³ CO₂ cochlea cochlear compartment compliance components computed tomography concentration constant counting rate cross-section cuff decrease detector diameter diastolic distribution effect elastin electrode energy equation filter fluid Fourier transform frame frequency function haemoglobin hair cells halothane heart rate impulse response increase left heart linear linear-attenuation coefficient lung manometer measured medical physics membrane method mmHg muscle normal obtained oxygen P₁ P₂ pacemaker patient peripheral resistance photons physicist physiological pulmonary pulsatile pulse ratio region Rendiconti S.I.F. sample scan scanner segment shown in fig shows signal stroke volume systolic techniques tissues transducer transmural pressure tube ultrasonic V₁ velocity venous system ventilation ventricle ventricular vessel viscoelastic volume wall wave form Windkessel X-ray zero